Device model: TG 209 F1 Libra
Manufacturer: NETZSCH-Gerätebau GmbH
Short description: Vacuum-tight thermo-microbalance for measurements up to 1100 °C with up to 200 K/min with water-cooled furnace. The thermostated ultra-microbalance offers a high resolution of 0.1 µg over the entire measuring range of 2000 mg. Samples with large mass changes can be analysed with high accuracy without switching measuring ranges. Additional information about endo- and exothermic processes (e.g. melting without mass loss or evaporation with mass loss) can be analysed via c-DTA signal.
Samples: Plastics, various fibre composites and composite materials
Applications: Material identification, analysis of material composition, investigation of thermal decomposition and degradation behaviour
Technical data | |
Measuring principle | Measurement of mass changes in defined atmospheres as a function of temperature and time |
Possible measuring environments | inert, oxidising, static, dynamic, under vacuum |
Heating and cooling rates | 0.001 K/min to 200 K/min |
emperature range | 10 to 1,100 °C |
Weighing range | up to 2,000 mg |
Signal resolution | 0.1 µg |
Sample crucible volume | up to 350 µl |
Cooling time 12 min | 1,100 bis 100 °C |
Other | Vacuum-tight construction up to 10-2 mbar (1 Pa), automatic sample changer (192 crucibles), coupling with gas analysis techniques such as FT-IR spectrometer and GC-MS possible |
Conformity to standards | DIN EN ISO 11358, ASTM D6370 |